Summary
Type: SPE
Application #: 301060
Catalog#: 730011 - SPE Cartridges, Chromabond, C18 ec, 1ml, PP, 100mg, 100/pk
Detection:
Conditions: Column conditioning:
3 column volumes methanol, then 3 column volumes 0.01 M potassium citrate
buffer pH 5.0
Sample application:
Slowly force or aspirate pretreated sample through the column
Column washing:
10 ml 0.01 M K-citrate buffer pH 5.0
Elution:
1 ml methanol / 0.01 M aqueous sodium hydrogen carbonate solution (1 : 1, v/v)
Authors: Macherey-Nagel
Source: LC Department